JavaScript seems to be disabled in your browser. For the best experience on our site, be sure to turn on Javascript in your browser.
Data Sheet (PDF)
The new OM-CP-VoltX Series consists of 4, 8, 12 and 16-channel, battery powered, DC voltage data logger with a reading rate of up to 4 Hz. X-Series devices ship with a standard USB-A to Micro USB cable for use with free OMCP PC based software. The X-series has a high memory capacity and increased download speed compared to previous generations of product.
This easy-to-use device features an 18 month typical battery life, user replaceable battery and programmable engineering units. The storage medium is non-volatile solid state memory, providing maximum data security even if the battery becomes discharged. To maximize memory capacity, users can enable or disable channels. For easy identification, each channel can be named with up to a ten digit title.
Using the OM-CP Data Logger Software, starting, stopping and downloading from the OM-CP-VoltX Series data loggers are simple and easy. Graphical, tabular and statistical data is provided for analysis and data can be viewed in multiple units, using the Engineering Units function. The data can also be automatically exported to Excel® for further calculations.
MEASUREMENT
Input Connection
Removable screw terminals
Model
Voltage Range
Voltage Resolution
Calibrated Accuracy
Input Impedance
Maximum Voltage Between Inputs to Ground
Analog Conversion Time
150 ms
Frequency Rejection
60+50 Hz
Temprature Coefficient
25 ppm/℃
Specified Accuracy Range
Nominal range @ 25 ℃
Engineering Units
Native Measurement units can be scaled to display measurement units of another type. This is useful when monitoring current outputs from different types of sensors such as temprature, CO2, flowrate and more.
GENERAL
Memory(All channels enabled)
4-channel: 698,709 readings per channel8-channel: 349,354 readings per channel12-channel: 232,903 readings per channel16-channel: 174,677 readings per channel
Start Modes
Software programmable immediate startor delay start, up to 6 months in advance
Real Time Recording
May be used with PC to monitor andrecord data in real time
LEDs
1 per channel and 2 status LEDs
Reading Rate
4 Hz up to 1 reading every 24 hours
Calibration
Digital calibration through software
Calibration Date
Automatically recorded within device
Battery Type
9 V lithium included, user replaceable
Battery Life
18 months typical
Data Format
Date and time stamped V, mV, µVengineering units specified throughsoftware
Time Accuracy
±1 minute/month
ComputerInterface
USB-A to micro USB cable (included);460,800 baud
OperatingSystemCompatibility
Windows XP SP3/7/8/10
Software Compatibility
Standard Software version 4.2.19.0 or later Secure Software version 4.2.18.0 or later
Operating Environment
-20 ℃ to +60 ℃(-4 ℉ to +140 ℉), 0 %RH to 95 %RH non-condensing
Dimensions
4-channel: 2.70 in × 7.25 in × 1.22 in(65.6 mm × 184.2 mm × 31.0 mm) 8-channel: 2.70 in × 7.25 in × 1.22 in(65.6 mm × 184.2 mm × 31.0 mm)12-channel: 2.70 in × 7.25 in × 1.68 in(65.6 mm × 184.2 mm × 42.7 mm)16-channel: 2.70 in × 7.25 in × 2.14 in(65.6 mm × 184.2 mm × 54.4 mm)
Weight
4-channel: 13 oz (368 g) 8-channel: 13 oz (368 g)12-channel:20 oz (580 g) 16-channel: 28 oz (800 g)
Enclosure
Black anodized aluminum
Approvals
CE, ROHS