- Description
Nominal size in mm
100, 160
Accuracy class
Model 532.51 NS 160: 0.6
Model 532.52: 1.0
Model 532.53: 1.6
Model 532.54: 2.5
The measuring accuracy is ensured for fluctuations in atmospheric pressure between 955 and 1,065 mbar (min. and max. of atmospheric pressure).
Scale ranges
0 ... 25 mbar to 0 ... 25 bar absolute pressure
Pressure limitation
Steady: Full scale value
Fluctuating: 0.9 x full scale value
Overpressure safety
Minimum 1 bar absolute pressure (atmospheric pressure), in addition 10 x full scale value, max. 25 bar absolute pressure
Permissible temperature
Ambient: -20 ... +60 °C
Medium: +100 °C maximum
Temperature effect
When the temperature of the measuring system deviates from the reference temperature (+20 °C):
max. ±0.8 %/10 K of full scale value
Ingress protection
IP 54 per EN 60529 / lEC 529